Submicron Aerosol Particle Losses in Metalized Bags

Alice Lecinski National Center for Atmospheric Research, Boulder, CO 80307

Search for other papers by Alice Lecinski in
Current site
Google Scholar
PubMed
Close
Restricted access

Abstract

Two new types of conducting bags were tested for aerosol particle storage and sampling, a 3M Company Velostat bag and a bag constructed from 3M Type 2100 Static Shielding Film. The half-lives of unipolar, unit-charged 0.025 μm, 0.050 μm and 0.090 μm sized aerosol particles stored in the Velostat bag and the film bag were 130, 190 and 270 min and 40, 70 and 180 min, respectively. These results depend upon the history of bag filling. The values given here apply to bags which had not previously been filled on the day of experimentation. The lifetimes exhibited by the aerosol particles stored in the Velostat bag are the longest found to data.

Abstract

Two new types of conducting bags were tested for aerosol particle storage and sampling, a 3M Company Velostat bag and a bag constructed from 3M Type 2100 Static Shielding Film. The half-lives of unipolar, unit-charged 0.025 μm, 0.050 μm and 0.090 μm sized aerosol particles stored in the Velostat bag and the film bag were 130, 190 and 270 min and 40, 70 and 180 min, respectively. These results depend upon the history of bag filling. The values given here apply to bags which had not previously been filled on the day of experimentation. The lifetimes exhibited by the aerosol particles stored in the Velostat bag are the longest found to data.

Save