1. Introduction
Water vapor is the most important natural greenhouse gas of the atmosphere and has a large impact on its radiative properties and hence on its thermodynamic balance. Despite the importance of water vapor in the climate system and weather forecasting, there is no technique available, neither ground-based nor spaceborne, that can provide continuous measurements of the humidity profile under all weather conditions. Ground-based microwave radiometers measuring the pressure-broadened emission line of water vapor at 22 GHz are good candidates to fill this gap.
Classical microwave profilers usually cover a frequency range from 20 to 30 GHz at a few individual channels for humidity measurements (Solheim et al. 1998; Crewell et al. 2001; Ware et al. 2003; Martin et al. 2006). The instrument presented here, however, was originally designed to measure water vapor profiles in the stratosphere and mesosphere as part of the Network for the Detection of Atmospheric Composition Change (NDACC; available online at http://www.ndacc.org). It covers a bandwidth of 1 GHz centered at 22.235 GHz with very high spectral resolution of 61 kHz, as required for observations in the middle atmosphere. Thus, only the peak of the H2O line is measured, as shown in Fig. 1. The spectral signature of clouds within the measured frequency interval of 1 GHz is in good approximation linear in frequency. The forward calculation in the retrieval is therefore based on a clear-sky atmosphere, and clouds are accounted for by adding an offset and a slope to the calculated spectrum. This approach allows the retrieval of water vapor profiles under almost all nonprecipitating weather conditions without additional information on clouds.
The paper is organized as follows: A short description of the instrument and the calibration method is presented in section 2. In section 3, the retrieval is discussed and the key characteristics are derived. A validation of the retrievals is presented in section 4.
2. Instrument and calibration
The emission spectra are measured by the Middle Atmospheric Water Vapor Radiometer (MIAWARA; Deuber et al. 2004), which is optimized for measuring the narrowband emission line from stratospheric and mesospheric water vapor. MIAWARA was developed and is operated by the University of Bern and is deployed close to Bern, Switzerland, at 47°N, 7°E at 900 m MSL. It is a single sideband heterodyne receiver converting the incoming signal at 22.235 GHz to an intermediate frequency of 0.5 GHz in two mixing steps. The signal is analyzed by a digital fast Fourier transform spectrometer with a spectral range from 0 to 1 GHz and a resolution of 61 kHz. Hence, the radiometer system is able to measure the emission from water vapor at frequencies between 21.735 and 22.735 GHz at very high resolution, as required for observations in the middle atmosphere. This high resolution, however, is of no advantage for the retrieval of tropospheric profiles, and for this study the high resolution raw spectra are binned into 20-MHz bins before being calibrated.


The thermal noise on a binned, single spectrum derived from a tipping curve is 0.2 K. For this study, bunches of 16 spectra have been summed up before the retrieval is performed to reduce the noise to 0.05 K. Because tipping curves are performed only every half an hour, the retrieved profiles are averaged over 8 h.
3. Retrieval of water vapor profiles


Within the spectral range of 1 GHz, the contribution from clouds to the measured brightness temperature can be reasonably well approximated with a linear term in the forward model. Thus, a polynomial of degree one is included in the forward model to account for clouds, and its coefficients (offset and slope) are part of the state vector x.


The random error (standard deviation of the differences between retrieved and true profile; see Fig. 2, right) is 5% of the mean value at the surface and increases linearly to 55% at 9 km. This corresponds to precisions given in Table 1, which compare well with the numbers reported by Crewell et al. (2001) for a 22-channel radiometer (10 water vapor channels).
We note that the retrieved offset and slope, which are expected to account for the effect of clouds, correlate well with the integrated liquid water. The correlation coefficients are r = 0.65 and r = 0.61 for the offset and the slope, respectively.
4. Results
Within 15 km distance to the site, where the radiometer is operated, 3–4 balloon soundings per week were performed during the time period from June 2007 to June 2008. The balloon payload was equipped with a Vaisala RS92 sensor. This dataset is used to evaluate the water vapor retrievals from the ground-based radiometer. An assessment of the accuracy of RS92 is presented by Miloshevich et al. (2009), and good agreement of nighttime measurements with the Cryogenic Frostpoint Hygrometer (CFH; Vömel et al. 2007), within 10% up to 5 km, is reported. However, a strong dry bias of daytime RS92 soundings compared to CFH of up to 30% at 10 km is reported. We corrected for this bias by applying the radiation correction suggested by Miloshevich et al. (2009) to the RS92 measurements, because all soundings considered in this study were performed during day. The correction is a function of pressure and solar zenith angle. Because clouds are expected to significantly reduce the radiation error, the correction was only applied to those portions of the RS92 humidity profiles that were lying above the uppermost cloud layers. Clouds are assumed in regions where the relative humidity is >95%.
The left panel of Fig. 3 shows the correlation coefficients between the retrievals, based on the measured spectra, and the coincident balloon soundings. Again, the seasonal cycle has been subtracted from the data for this analysis. The correlation matrix is in good agreement with the one derived from the simulation (Fig. 2). The scatterplots of the retrievals and RS92 measurements are shown in the right panels of Fig. 3. The performance of the retrieval is good up to 5 km, with the restriction that high values are slightly overestimated. At 6 and 7 km, the correlation is degraded and a wet bias is noted. At 8 and 9 km, the retrieval is not able to get the high values right at this altitude because of the increasing influence of the a priori profile. As mentioned in section 3, including a seasonal cycle in the a priori profile would improve the performance at these levels.
Figure 4 shows time series of water vapor measured by RS92 and by the radiometer at four pressure levels. The features discussed earlier are evident, namely the high correlation between the two datasets and the wet bias of the microwave retrievals compared to RS92. Visual inspection of the time series reveals that the agreement is generally better after September 2007 (most pronounced at 420 hPa). In September 2007, the radiometer experienced a few hardware changes, including the installation of a blade to protect the hot load (ambient load) from direct insolation. This leads to a better definition of the temperature of the hot load, and thus improves the calibration. We are not able to quantify the effect of these changes on the retrievals, but we note that they improved the calibration and are likely to explain the better performance after September 2007.
In the following analysis, we consider only data after September 2007, which reduces the total of 208 profile pairs to a total of 145. The mean of the differences between the radiometer and RS92 as a function of altitude is presented in the left panel of Fig. 5. A wet bias of 10%–20% is apparent at altitudes below 6 km. The random error (standard deviation of the differences between radiometer and RS92) is shown in the right panel of Fig. 5, along with the natural variability (dashed line) and the random error of the a priori profile (dashed–dotted line). Because the a priori profile is set to the measured surface value at the ground, it performs slightly better than the retrieval up to 2 km. Above 2 km, information from the measurement is added to the a priori profile and the retrieval performs well up to 9 km, showing a bump in the random error profile reflecting decreased correlation with RS92 at 6–7 km. The random error of the retrievals is in good agreement with the one derived from the simulation.
The temperature profiles from ECMWF are on average 1.5 K warmer than the temperatures measured by the radiosonde. Using the temperature profiles from the radiosonde in the retrieval increases, the bias in water vapor of the microwave retrievals compared to RS92 by 1%–2%.
5. Summary and conclusions
We presented a characterization and first results of the retrieval of tropospheric water vapor profiles from pressure-broadened emission spectra. The measured spectra have a bandwidth of 1 GHz at a resolution of 20 MHz and are centered at 22.235 GHz. This is in contrast to conventional microwave profilers, which cover a broader spectral range at a much coarser resolution. The spectra are calibrated by means of a tipping curve measurement and inverted with an optimal estimation algorithm, because the small-bandwidth clouds can be accounted for with an offset and a slope in the forward model and no additional information on clouds is needed. This is a major advantage of this approach, because water vapor profiles of good quality can be retrieved under almost all nonprecipitating weather conditions.
A simulation of the retrieval demonstrates the potential of this experimental setup and shows that measurements are in principle possible up to 9 km with a vertical resolution of ≈5 km. This performance is comparable to conventional microwave profilers. The real measurements have been evaluated with 200 coincident balloon soundings. The RS92 data have been corrected for the radiation dry bias. The performance of the real retrieval is good throughout the whole troposphere, showing a correlation with RS92 better than 0.7. Best performance is found up to the middle troposphere (5 km). The random error derived from the measurements is 10% at the surface and increases to 55% at 9 km. In general, the retrieved profiles show a wet bias compared to RS92 of 10%–20% between 2 and 5 km. This wet bias is substantial and needs to be investigated further. We note that, in the retrieval, water vapor is assumed to be normally distributed. This assumption is questionable and could introduce a bias. However, we conclude that this unconventional setup is well suited to measure tropospheric humidity up to 9 km, independent from additional cloud information under a large range of weather conditions.
This work has been supported by the Swiss National Science foundation under Grant 200020-115882/1 as well as through the project SHOMING financed by MeteoSwiss within GAW. We acknowledge the support of the European Commission through the GEOMON Integrated Project under the 6th Framework Program (Contract FOP6-2005-Global-4-036677).
REFERENCES
Buehler, S. A., , Eriksson P. , , Kuhn T. , , von Engeln A. , , and Verdes C. , 2005: ARTS, the atmospheric radiative transfer simulator. J. Quant. Spectrosc. Radiat. Transf., 91 , 65–93. doi:10.1016/j.jqsrt.2004.05.051.
Crewell, S., , Czekala H. , , Löhnert U. , , Simmer C. , , Rose T. , , Zimmermann R. , , and Zimmermann R. , 2001: Microwave Radiometer for Cloud Carthography: A 22-channel ground-based microwave radiometer for atmospheric research. Radio Sci., 36 , 621–638.
Deuber, B., , Kämpfer N. , , and Feist D. , 2004: A new 22-GHz radiometer for middle atmospheric water vapour profile measurements. IEEE Trans. Geosci. Remote Sens., 42 , 974–984.
Eriksson, P., , Jiménez C. , , and Buehler S. A. , 2005: Qpack, a general tool for instrument simulation and retrieval work. J. Quant. Spectrosc. Radiat. Transf., 91 , 47–64. doi:10.1016/j.jqsrt.2004.05.050.
Han, Y., , and Westwater E. R. , 2000: Analysis and improvement of tipping calibration for ground-based microwave radiometers. IEEE Trans. Geosci. Remote Sens., 38 , 1260–1276.
Martin, L., , Schneebeli M. , , and Mätzler C. , 2006: Tropospheric water and temperature retrieval for ASMUWARA. Meteor. Z., 15 , 37–44. doi:10.1127/0941-2948/2006/0093.
Miloshevich, L. M., , Vömel H. , , Whiteman D. N. , , and Leblanc T. , 2009: Accuracy assessment and correction of Vaisala RS92 radiosonde water vapor measurements. J. Geophys. Res., 114 , D11305. doi:10.1029/2008JD011565.
Rodgers, C. D., 2000: Inverse Methods for Atmospheric Soundings. World Scientific, 240 pp.
Rosenkranz, P. W., 1998: Water vapor microwave continuum absorption: A comparison of measurements and models. Radio Sci., 33 , 919–928.
Solheim, F., , Godwin J. R. , , Westwater E. R. , , Han Y. , , Keihm S. J. , , Marsh K. , , and Ware R. , 1998: Radiometric profiling of temperature, water vapor and cloud liquid water using various inversion methods. Radio Sci., 33 , 393–404.
Vömel, H., , David D. , , and Smith K. , 2007: Accuracy of tropospheric and stratospheric water vapor measurements by the cryogenic frost point hygrometer: Instrumental details and observations. J. Geophys. Res., 112 , D08305. doi:10.1029/2006JD007224.
Ware, R., , Carpenter R. , , Güldner J. , , Liljegren J. , , Nehkorn T. , , Solheim F. , , and Vandenberghe F. , 2003: A multichannel radiometric profiler of temperature, humidity, and cloud liquid. Radio Sci., 38 , 8079. doi:10.1029/2002RS002856.

Absorption coefficient between 10 and 40 GHz at 850 hPa for water vapor, cloud liquid water, and oxygen. The shaded area marks the bandwidth of the radiometer.
Citation: Journal of Atmospheric and Oceanic Technology 27, 1; 10.1175/2009JTECHA1334.1

Analysis of the simulated retrievals. (left) Matrix of correlation coefficients between the true and the retrieved profiles. Because the true profiles are based on RS92 measurements, they are only taken into account up to 11.5 km. (right) Scatterplots of the true and the retrieved data, where r is the correlation coefficient.
Citation: Journal of Atmospheric and Oceanic Technology 27, 1; 10.1175/2009JTECHA1334.1

Analysis of the real retrievals. (left) Matrix of correlation coefficients between RS92 and the microwave retrievals (MIAWARA). RS92 measurements are only taken into account up to 11.5 km. (right) Scatterplots of RS92 and radiometer data, where r is the correlation coefficient. RS92 profiles have been corrected for the radiation dry bias (see section 4).
Citation: Journal of Atmospheric and Oceanic Technology 27, 1; 10.1175/2009JTECHA1334.1

Time series of H2O mixing ratio as measured by RS92 (gray lines) and the ground-based microwave radiometer (black lines) at four different pressure levels. RS92 profiles have been corrected for the radiation dry bias (see section 4).
Citation: Journal of Atmospheric and Oceanic Technology 27, 1; 10.1175/2009JTECHA1334.1

(left) Mean and (right) std dev of the relative differences between the radiometer (MIAWARA) and RS92 profiles considering only measurements after 1 Oct 2007 (solid black line; 145 profiles) and considering all measurements (solid gray line; 203 profiles). (right) The natural variability of water vapor (dashed line) and the random error of the a priori (dashed–dotted line) are also displayed.
Citation: Journal of Atmospheric and Oceanic Technology 27, 1; 10.1175/2009JTECHA1334.1
Random error (std dev of differences between retrieved and true profiles) of the water vapor retrievals derived from the simulation (VMR is the volume mixing ratio).