1. Introduction



Cross sensitivity is defined here as an observed change in CO2 density caused by a change in water vapor density. In this study we consider cross sensitivity to water vapor after processing CO2 fluxes to account for IR density variations. Temperature sensitivity related to CO2 flux retrieval is not investigated here. Open ocean studies have shown that measuring CO2 density when water vapor density is changing in the sampling air can give erroneous CO2 measurements as a result of cross sensitivity (Kohsiek 2000; Kondo et al. 2014; Landwehr et al. 2014; Blomquist et al. 2014).
According to a study by Kohsiek (2000) in which two gas analyzers were compared (NOAA and KNMI sensors), the two most likely explanations for the cross-sensitivity problem are the pressure band broadening effect and the effect of the presence of liquid water in the optical path. Commercially used sensors, such as LI-7200 and LI-7500 (from Li-Cor Inc., Lincoln, Nebraska) have, according to the manufacturers, corrections for the pressure broadening effect incorporated into their software. However, according to Kondo et al. (2014), the corrections introduced by the manufacturers were not enough and consequently the instruments could overestimate the CO2 mixing ratio and the magnitude of the flux.
It is notable that all studies showing high-cross-sensitivity errors in the CO2 fluxes are open ocean or coastal sea studies. Even after following the calculations of Webb et al. (1980), the CO2 fluxes from open-path and enclosed-path instruments can be an order of magnitude higher than expected values based on typically accepted bulk flux parameterizations (Prytherch et al. 2010; Landwehr et al. 2014). Landwehr et al. (2014) showed that a postprocessing correction method, which was suggested by Prytherch et al. (2010) [Peter K. Taylor (PKT) corrections], could not be used to correct the CO2 fluxes and instead supported the conclusions of Miller et al. (2010) that an order-of-magnitude bias in the measured CO2 fluxes could be removed only if the sample air is dried. Landwehr et al. (2014) found that the CO2 flux estimates from the dried and undried Li-Cor sensors agreed for very low ambient latent heat fluxes (<7 W m−2). The terminology of dried sensor for both Landwehr et al. (2014) and this study means that the intake sample airstream has been passed through a Nafion dryer that effectively removed fluctuations of water vapor before the airstream is passed to the gas analyzer. The practical recommendation of Miller et al. (2010), Landwehr et al. (2014), and Blomquist et al. (2014) would be to use closed or enclosed-path instruments with a diffusion dryer to remove water vapor fluctuations for EC flux measurements of CO2 over the open ocean.
Blomquist et al. (2014) showed convincingly that the measured mean CO2 concentration can be several parts per million (ppm) different if a Nafion air dryer is used. From the studies by Kondo et al. (2014) and Blomquist et al. (2014), it is less clear to what extent the actual CO2 fluxes are changing significantly by drying the air. Neither of these studies focused specifically on the fluxes, although Blomquist et al. (2014) discussed, based on their analysis, that errors in the CO2 flux of about 10% of the specific humidity may be possible. By recomputing CO2 from the raw absorbances with an adjustment of a water crosstalk constant (aw = 1.7 instead of 1.15), Blomquist et al. (2014) showed a large removed offset in measured CO2 concentration in moist air. However, (Blomquist et al. 2014, p. 261) also commented that “recomputing fluxes for the period without the dryer from the corrected time series did not yield a significant improvement in flux variance and bias.” A study more directly focusing on the effects of drying the air and how it affects measured CO2 fluxes rather than the level of CO2 concentration is therefore complementary to existing published literature.
Spectral attenuation or high-frequency flux loss is an additional issue for all eddy covariance measurements because they are bandwidth limited (Blomquist et al. 2014). Resolution of the smallest eddies (highest frequencies) is limited by sensor separation, sampling frequency, and low-pass filtering from tubing in closed and enclosed-path systems. Correction for this effect is not the focus here but two general approaches to perform spectral attenuation corrections are using transfer functions (Moore 1986; Horst 1997; Massman 2000) and spectral similarity methods (Hicks and McMillen 1988). The spectral attenuation corrections are better described and are not as large as the magnitude order errors discussed for oceanic flux experiments.
An additional concern for the ocean flux community is the presence of hygroscopic deposits on the optics of CO2 gas analyzers, such as sea salt, which could also affect the measurements (Kondo et al. 2014). Blomquist et al. (2014) discussed that if hygroscopic contamination is a major factor inducing cross sensitivity, then the magnitude of cross correlation between water vapor and CO2 should decrease following a wash cycle on their LI-7500 analyzer, which they did not observe. They therefore argued that optical contamination may not have been the source of additional cross correlation in their case, or perhaps that the relevant contaminants are resistant to removal by rinsing. It is also possible that the saltwater films on the sensor lenses recover too fast after washing (within minutes) to allow the observation of a “clean” flux signal, which is typically computed over time scales of 10 min to 1 h. In the context of determining an appropriate eddy covariance flux that relates to the dynamic and turbulent scales of the atmospheric boundary layer, it is appropriate to use stationarity tests. A common approach to determine a nonfixed averaging time includes using ogive curves to assure that extending the cross covariance integral does not add significantly more than 95% for the flux within each chosen averaging period.
A third concern in the flux community for both land and ocean studies is that when using open-path sensors, surface heating from the instrument itself could lead to warming of the air in the measuring path and to errors in the CO2 fluxes, which will not be corrected by the density corrections because normally the temperature used in the density correction is not measured directly in the measuring path of the gas analyzer (Burba et al. 2008). The third issue is substantial only in very cold climates. Wang et al. (2016) compared spectroscopic temperature sensitivity to the self-heating issue and found no evidence for self-heating as the cause of a probable uptake CO2 flux artifact. The temperature range studied in Wang et al. (2016) was from about −20° to +25°C. Our study will provide field measurement results on the former issues of water vapor and sea salt. In the case of measurements on ships, there are several additional concerns related to ship motion and airflow distortion not further addressed here—see instead for instance Blomquist et al. (2014), Prytherch et al. (2015), and Landwehr et al. (2015).
Here we aim to investigate possible disturbances by water vapor and salt contamination on turbulent fluxes and respond to two main questions, Does drying the air influence measured CO2 fluxes, and will measured CO2 fluxes be affected by sea salt contamination on the windows of instruments? We discuss some results as they are presented in section 3 and also summarize our recommendations for measuring turbulent fluxes of CO2 with infrared gas analyzers in our summary and conclusions sections.
2. Sites, measurements, and data processing
Field measurements were conducted at an agricultural site, Marsta, located close to Uppsala, Sweden (Halldin et al. 1999). The period from April to June 2015 is used as the main dataset in this study. This is complemented by a more limited set of measurements from the Östergarnsholm marine field site located on a small island in the Baltic Sea. These measurements were conducted during the summer of 2015 (July–September).
a. Field measurements from Marsta
In Marsta two open-path Li-Cor 7500 gas analyzers and two enclosed-path Li-Cor 7200 instruments were mounted next to one Gill WindMaster sonic anemometer at a measurement height of 7 m (see Fig. 1). To minimize the loss of measured high-frequency fluctuations, the length of the intake tubes was kept at about 0.5 m. The inner diameter of the tubes was 5 mm and a flow rate of 18 L min−1 was used. The distance between the instruments were kept as short as possible (maximum 40 cm), and the scalar sensors were placed slightly below the anemometer, consistent with recommendations to minimize flux loss caused by sensor displacement (Horst and Lenschow 2009; Nilsson et al. 2010). The agricultural landscape surrounding the site is very flat, and in some wind direction sectors the upwind fetch is undisturbed crop fields for 4–5 km.

A picture of the instrument setup with four Li-Cor instruments and the Gill sonic anemometer at the field measurement site Marsta near Uppsala.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

A picture of the instrument setup with four Li-Cor instruments and the Gill sonic anemometer at the field measurement site Marsta near Uppsala.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
A picture of the instrument setup with four Li-Cor instruments and the Gill sonic anemometer at the field measurement site Marsta near Uppsala.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
There was very low vegetation for most of the period as it was in the beginning of the growing season. Local farmers nearby noted, however, an unusually early start of their spring activities in 2015 (with fertilization of the fields beginning around 12 April and spring seeding completed about 2–3 weeks later). This was mainly related to April being very dry (only about 10 mm were recorded, which is about one-third of the normal precipitation) and warmer than normal (6.8°C compared to 4.0°C for the climatic period of 1961–90). In May 2015 in contrast the mean temperature was close to normal (9.6°C compared to 10.2°C as a mean value for the period 1961–1990) and there were much more precipitation, about 85 mm—almost 3 times more than normal. These different conditions are mentioned here because data from these two months were used to determine a reasonable selection criteria in our quality control of CO2 fluxes. In June 2015 we had 16 days with precipitation, but only 1 day with more than 10 mm. The monthly precipitation was only marginally less (1 mm) than the climatic mean value of 45 mm for the period 1961–90, and the monthly mean temperature was 14°C, which is 1°C cooler than normal.
The field measurements from Marsta were divided into four periods, which correspond to
Measurements in April, when all four Li-Cor instruments were used in a reference setup, with no drying of intake air and no influence from an external source of salt.
Measurements in May, when one of the enclosed-path Li-Cor 7200 instruments (called Li72b) had a Nafion membrane dryer (PD-200T) mounted to remove water vapor fluctuations. For more details on the application of this dryer system in the context of marine measurements of CO2 fluxes, refer to Landwehr et al. (2014).
Measurements in June, which had the same setup as in period 2, which is used as a reference for the sea salt experimental field test.
Measurements in June, with a smaller amount of data for which the enclosed-path Li-Cor 7200 instruments (with and without the Nafion dryer) was sprayed with a saline solution to mimic the effect of sea salt contamination on the instrument windows.
This division of data into these four periods—periods 1–4, respectively—was done to perform tests of the separate effects of drying the air, to investigate the sea salt contamination issue, and to compare potential differences in measured fluxes in the reference setup(s).
b. Field measurements from Östergarnsholm
Field measurements were also conducted at the Swedish marine Integrated Carbon Observation System (ICOS) site Östergarnsholm. This site is located on the island Östergarnsholm, 4 km from the eastern coast of the larger island Gotland in the Baltic Sea. The southern part of the island is very flat and rises only a couple of meters above sea level. A 30-m instrumented tower is located here (see Fig. 2), which has been used for measuring CO2 fluxes and atmospheric turbulence and mean parameters semicontinuously for the past 20 years (Rutgersson et al. 2008; Högström et al. 2008).

A picture of the 30-m tower at the Östergarnsholm site with the 10-m level indicated. Also seen is a smaller tower instrumented for measurements of marine aerosols.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

A picture of the 30-m tower at the Östergarnsholm site with the 10-m level indicated. Also seen is a smaller tower instrumented for measurements of marine aerosols.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
A picture of the 30-m tower at the Östergarnsholm site with the 10-m level indicated. Also seen is a smaller tower instrumented for measurements of marine aerosols.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
During the period between July and September 2015, two Li-Cor 7200 instruments and one Li-Cor 7500 instrument were mounted close to a CSAT sonic anemometer at a nominal measurement height of 10 m MSL. Only data with wind from the 80°–220° sector were used to ensure the measured fluxes represent sea conditions.
The instrumental setup was essentially identical to that of period 2 in Marsta but only one Li-Cor 7500 instrument was used. One of the Li-Cor 7200 instruments had a Nafion dryer installed to remove water vapor fluctuations, whereas the other Li-Cor 7200 was left in a standard operating mode measuring both CO2 fluxes and latent heat fluxes.
c. Overview of encountered environmental conditions and CO2 fluxes at the two sites
To make a brief comparison of the conditions at the two sites, we list in Table 1 the mean, maximum, and minimum values of wind speed, temperature, and humidity, as well as quality-controlled fluxes of CO2. If we compare the conditions at the two sites during the time of deployments, we had larger diurnal cycles at Marsta compared to Östergarnsholm. The lowest temperature in Marsta of −0.3°C also came in early spring and the highest value of 27.8°C was in June. At Östergarnsholm the observed temperature range was not as large, ranging from 12.8° to 23.2°C, and had a mean value of 18.1°C, which can be compared with the mean value at Marsta, which was 12.9°C. The wind speed was higher for Östergarnsholm with a mean value of 6.8 compared with 2.5 m s−1 at Marsta, and the highest hourly wind speed was 13.7 m s−1 at Östergarnsholm, whereas it was 8.5 m s−1 at Marsta. This is a general observation for the coastal site compared to the inland site, that it is windier at Östergarnsholm. Higher humidity is also generally found at the marine site compared to the inland site with differences of 3.7 g kg−1 in the mean for the periods considered. In fact, the maximum value at Marsta for the considered periods was 9.0 g kg−1, which is comparable to the mean value at Östergarnsholm, which was 8.9 g kg−1.
Mean, maximum, and minimum of meteorological parameters and the flux of CO2 (two different units) at Marsta during April–June 2015 and at Östergarnsholm during July–September 2015.


When it comes to the range of encountered CO2 fluxes, the values reported in Table 1 refer to fluxes that have undergone visual inspection and quality control in a number of steps, which will be described in more detail later in this paper. Here, however, we first want to give an overview of the magnitudes encountered at these two sites, which are indeed very different from each other. The mean flux at the inland agricultural site was −2.1 µmol m−2 s−1, which is more than 13 times larger than the mean flux at the marine site. The range of observed fluxes at Marsta was from 7.2 to −23.0 µmol m−2 s−1, which is quite comparable to the ranges discussed for instance in Baldocchi et al. (2001) when comparing an Italian CarboEuroFlux site with the AmeriFlux reference (their Fig. 2). For Östergarnsholm the mean value of −0.16 µmol m−2 s−1 or about −5.0 mol m−2 yr−1 corresponds well to previous estimates of air–sea exchange of CO2 at the site for the month of August presented by Rutgersson et al. (2009). A smaller flux and a range of fluxes between 0.5 and −0.79 µmol m−2 s−1 was thus observed at Östergarnsholm compared to Marsta. For more detailed discussions on estimates of CO2 exchange in coastal zones and inland areas, refer to Grachev et al. (2011), which showed similar magnitudes of averaged CO2 turbulent flux over the Gulf of Mexico of −7.47 mol m−2 yr−1 but very different results over suburban areas. Different results with an average CO2 flux from the ocean to the atmosphere of 3.9 mol m−2 yr−1 has also been reported for open ocean conditions in the equatorial Pacific (McGillis et al. 2004). Different marine areas show different results also depending on the season, but the magnitude of fluxes are often lower than at many sites over land.
In Table 1 we have listed CO2 fluxes in both μmol m−2 s−1 and ppm m s−1 for the convenience of readers used to specific units. For the instrument comparison we choose to use ppm m s−1, as it is a convenient unit to use when studying kinematic CO2 flux when vertical wind speed is measured in meters per second and CO2 is in parts per million. Positive flux values denote upward fluxes (CO2 respiration), and negative values correspond to downward fluxes (CO2 photosynthesis).
d. Data processing




The choice of 10 min as an averaging period was for quality-controlled data, which were also checked as corresponding to an approximate range of frequencies between 1/1000 and 3/1000 Hz, where the flux contribution was smaller compared to higher and lower frequencies. Such a cospectral “gap” separated what we considered to be the turbulent flux range from more low-frequency correlations. The choice of the 10-min averaging time is used here to stabilize the observations of CO2 fluxes rather than calculating a final flux value that can be attributed to specific surface characteristics. Also, the spectral slope of the vertical wind components was checked, and the overall behavior of CO2 spectra and cospectra from all the instruments was checked manually. No frequency corrections for loss of variance or flux were performed, as this was not the focus of the current investigation. We consider the approach taken as sufficient to evaluate the effect of drying and the effect of sea salt on CO2 fluxes. Additional processing may be necessary if attribution of measured fluxes to specific surfaces are desirable. In such a case, further review of data processing details may be needed (Lee et al. 2005; Nakai et al. 2006; Starkenburg et al. 2016), including flux footprint modeling (Kljun et al. 2004; Vesala et al. 2008).
3. Results and discussion
We present our results in three subsections. To meet an acceptable data quality, we selected thresholds on the quality indicators and this procedure is presented first. Second, we present the results from field tests of drying the air to remove water vapor fluctuations. These include data from both Marsta, the agricultural site, and Östergarnsholm, the marine site. Finally, we present results about sea salt contamination.
a. Quality control
The gas analyzers give a signal strength index called relative signal strength indicator (RSSI), which was saved for all our instruments at high frequency (20 Hz). This can be used either as a quality control measure by itself or used to form additional quality control parameters. We noted that the RSSI value itself could differ systematically between instruments, which is typical between instrument brands. Sometimes they differed by a near constant under very good conditions with no rain or other disturbances. Therefore, from 10-min mean values of RSSI it was difficult to choose a specific value that would separate the high-quality data from the lower-quality data and at the same time not remove too much data (see the appendix). Instead we choose the variance of RSSI (
The squared correlation coefficient for the CO2 flux (black lines) from the two Li-Cor 7500 instruments in Fig. 3a and for the two Li-Cor 7200 instruments in Fig. 3b is shown as a function of different threshold values chosen for the variance of the RSSI value. Also shown is the fraction of remaining data using different threshold values (blue) and the threshold value

The squared correlation coefficient for the CO2 flux (
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

The squared correlation coefficient for the CO2 flux (
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
The squared correlation coefficient for the CO2 flux (
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
In May with many rain events (it rained 22 out of 31 days) and with the introduction of the Nafion dryer on one of the Li-Cor 7200 instruments, the squared correlation coefficient for the CO2 fluxes from the two Li-Cor 7200 instruments dropped to low values (far below 0.8) when a less strict threshold value was chosen. Similar results were also found for data in June. The appendix shows (Fig. A1) that the drop in the correlation coefficient occurred between the dried Li-Cor 7200b and Li-Cor 7500 instruments but not between the enclosed-path sensors. This indicates the risk in accepting data from dried sensors with a threshold limit set higher than 0.001. Selecting a good quality control parameter and threshold value is of course not trivial, and using the appendix (Figs. A2–A4) we discuss and compare our choices to another criteria based on instead choosing a limit for RSSI values scaled with their median RSSI value (
From the choice of a threshold limit of
b. Does drying the air influence measured CO2 fluxes?
1) Marsta
From period 1 and the field experiments in Marsta, there were 2044 high-quality data (10-min values) remaining with our chosen threshold value of
The latent heat fluxes from the two Li-Cor 7200 instruments (Li72a and Li72b) and one of the Li-Cor 7500 instruments (Li75a) are compared (Fig. 4a) to the latent heat flux measured by the other Li-Cor 7500 instrument (Li75b) for period 2. The comparison of the two Li-Cor 7500 instruments (black crosses) shows systematic differences of less than 2%. This reflects instrumental uncertainty and a different sampling of the smallest-scale scalar fluctuations by the different sensors.

A comparison of measured (a) latent heat flux and (b) CO2 flux between the Li75b instrument and the other three instruments for period 2 when a Nafion dryer was installed on one of the Li-Cor 7200 instruments (Li72b). A comparison of the two Li-7500 instruments is shown (black crosses). A comparison of fluxes between the undried Li-7200 without a mounted Nafion dryer and one of the Li-7500 instruments is shown (blue circles). A comparison of the CO2 fluxes with the dried Li-7200 to the same Li-7500 (red dots). Three fitted lines are included with the data in the corresponding color scheme, and the fitted expressions are given in the upper left of each plot.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

A comparison of measured (a) latent heat flux and (b) CO2 flux between the Li75b instrument and the other three instruments for period 2 when a Nafion dryer was installed on one of the Li-Cor 7200 instruments (Li72b). A comparison of the two Li-7500 instruments is shown (black crosses). A comparison of fluxes between the undried Li-7200 without a mounted Nafion dryer and one of the Li-7500 instruments is shown (blue circles). A comparison of the CO2 fluxes with the dried Li-7200 to the same Li-7500 (red dots). Three fitted lines are included with the data in the corresponding color scheme, and the fitted expressions are given in the upper left of each plot.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
A comparison of measured (a) latent heat flux and (b) CO2 flux between the Li75b instrument and the other three instruments for period 2 when a Nafion dryer was installed on one of the Li-Cor 7200 instruments (Li72b). A comparison of the two Li-7500 instruments is shown (black crosses). A comparison of fluxes between the undried Li-7200 without a mounted Nafion dryer and one of the Li-7500 instruments is shown (blue circles). A comparison of the CO2 fluxes with the dried Li-7200 to the same Li-7500 (red dots). Three fitted lines are included with the data in the corresponding color scheme, and the fitted expressions are given in the upper left of each plot.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
A comparison of the undried enclosed-path Li-7200 instrument (Li72a) to the open-path Li75b (blue circles) showed losses of latent heat flux for this period of about 9% when using enclosed-path systems (similar percentage losses were found for both Li72a and Li72b for period 1). This loss of flux mainly occurred for frequencies above about 0.3 Hz according to cospectral analysis (not shown here). This loss of flux likely results from damping of fluctuations in tubing. The comparison of the latent heat fluxes from the dried Li72b instrument to the Li75b latent fluxes (red dots) shows that the Nafion dryer worked well to reduce the latent heat fluxes to nearly 0 (less than 1% of the Li75b fluxes in the mean).
From linear fitted expressions of the corresponding measured CO2 fluxes from period 2 (Fig. 4b) it is first noted that there was essentially no difference in CO2 flux between the two Li-Cor 7500 instruments (less than 1% systematic difference). Comparing the two Li-Cor 7200 instruments to one of the Li-Cor 7500 instruments shows that both the dried and undried sensors have a systematic loss in measured CO2 flux of about 6%–7%. Thus, the dried CO2 signal (red dots) and undried CO2 signal (blue circles) agree equally well with the open-path system. For the CO2 flux, a cospectral analysis showed that losses of flux took place over a large range of frequencies above 3/1000 Hz. Most notable was, however, that for the Li72b instrument, there was no mean flux contribution for frequencies above 0.7 Hz, where also spectral variance of the CO2 signal was substantially reduced for enclosed-path instruments compared to open-path instruments. Some of the flux loss could be corrected using several of the approaches previously mentioned (Moore 1986; Hicks and McMillen 1988; Horst 1997; Massman 2000), but this would not affect our main results concerning the effects of drying or the effects from sea salt on the windows of the instruments.
The difference between the two enclosed-path sensors is very small. In Fig. 5, the measured CO2 fluxes from the dried Li72b are directly compared to the fluxes from the undried Li72a instrument. This reveals that the two measured fluxes (dried and undried) agree very well, with a systematic difference of less than 2%. Only a few 10-min flux values stand out from the rest. Attribution of the cause of these data would require further analysis, but it does not influence our main results in any way.

A direct comparison of measured CO2 flux from the undried Li-Cor 7200a instrument and the Li-Cor 7200b instrument with a Nafion dryer during period 2. The data (black crosses) and a linear fit to the data (black line) are shown. The fitted expression is also given in the upper-left corner.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

A direct comparison of measured CO2 flux from the undried Li-Cor 7200a instrument and the Li-Cor 7200b instrument with a Nafion dryer during period 2. The data (black crosses) and a linear fit to the data (black line) are shown. The fitted expression is also given in the upper-left corner.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
A direct comparison of measured CO2 flux from the undried Li-Cor 7200a instrument and the Li-Cor 7200b instrument with a Nafion dryer during period 2. The data (black crosses) and a linear fit to the data (black line) are shown. The fitted expression is also given in the upper-left corner.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
Box plots of the difference in measured CO2 flux between the Li75b instrument and the other three instruments for periods 1 and 2 corroborate the results of small changes between periods (Fig. 6). The light gray boxes on the left show the comparison of the two Li-Cor 7500 instruments with no real change of median value in between the two periods but a slightly larger spread of the variations for period 2. Similar results are shown for the differences between the Li72a and Li75b fluxes (white boxes) and for Li72b and Li75b flux differences (dark gray boxes). The increase in spread of the differences in period 2 is not exactly understood.

Boxplots of the difference in measured CO2 flux between the Li75b instrument and the other three instruments for period 1 (p.1) and period 2 (p.2). In period 2 a Nafion dryer was mounted on one of the Li-7200 instruments (Li-72b). The light gray boxes show the comparison of the two Li-7500 instruments. The white boxes show the comparison of the Li-7200 instrument with a Nafion dryer installed during period 2. The dark gray boxes show the comparison of the Li-7200 instrument without a Nafion dryer (Li72a). The median value is marked (dash–dotted line), and the boxes mark the 25th and 75th percentiles. Outliers are drawn as plus (+) signs if they are larger than Q3 + 1.5(Q3 − Q1) or smaller than Q1 − 1.5(Q3 − Q1), which define the whiskers of the data. Here Q1 and Q3 are the 25th and 75th percentiles, respectively.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

Boxplots of the difference in measured CO2 flux between the Li75b instrument and the other three instruments for period 1 (p.1) and period 2 (p.2). In period 2 a Nafion dryer was mounted on one of the Li-7200 instruments (Li-72b). The light gray boxes show the comparison of the two Li-7500 instruments. The white boxes show the comparison of the Li-7200 instrument with a Nafion dryer installed during period 2. The dark gray boxes show the comparison of the Li-7200 instrument without a Nafion dryer (Li72a). The median value is marked (dash–dotted line), and the boxes mark the 25th and 75th percentiles. Outliers are drawn as plus (+) signs if they are larger than Q3 + 1.5(Q3 − Q1) or smaller than Q1 − 1.5(Q3 − Q1), which define the whiskers of the data. Here Q1 and Q3 are the 25th and 75th percentiles, respectively.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
Boxplots of the difference in measured CO2 flux between the Li75b instrument and the other three instruments for period 1 (p.1) and period 2 (p.2). In period 2 a Nafion dryer was mounted on one of the Li-7200 instruments (Li-72b). The light gray boxes show the comparison of the two Li-7500 instruments. The white boxes show the comparison of the Li-7200 instrument with a Nafion dryer installed during period 2. The dark gray boxes show the comparison of the Li-7200 instrument without a Nafion dryer (Li72a). The median value is marked (dash–dotted line), and the boxes mark the 25th and 75th percentiles. Outliers are drawn as plus (+) signs if they are larger than Q3 + 1.5(Q3 − Q1) or smaller than Q1 − 1.5(Q3 − Q1), which define the whiskers of the data. Here Q1 and Q3 are the 25th and 75th percentiles, respectively.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
Nevertheless, the systematic differences between instruments are relatively small in terms of differences between instruments (%), and the residual difference between the Li75b fluxes and the measured fluxes from the other three instruments is closer to being Gaussian than the CO2 fluxes themselves from any single instrument. We can therefore rely on standard two-sample t tests for the change of mean values of these difference time series.
The mean CO2 flux from period 1 is about −0.005 and about −0.067 ppm m s−1 for period 2 (Table 2). It is reported in columns 2 and 3 (along with the mean difference after subtracting the Li75b flux). The change in mean for the difference time series between period 1 and period 2 is also given in column 4. Because of the high number of available data (more than 1000 10-min values) these rather small changes were found to be statistically significant at more than a 95% confidence level according to the t tests, but the largest change observed is for the Li72a instrument without a Nafion dryer. This change is also only about 50% larger than the difference noted between the two Li-Cor 7500 instruments; therefore, it can be considered small. The change for the Li72b instrument with a mounted Nafion dryer in period 2 was even smaller. Therefore, we can conclude that drying the sample air does not alter the CO2 signal very much.
Statistical summary of the experimental field test with and without a Nafion dryer mounted on one of the Li-7200 instruments in period 2. From period 1 in April 2015, 2044 high-quality data (10-min values) were available and 1381 high-quality data were available from period 2 in May 2015 when the Nafion dryer was mounted on Li-7200b. Here change is defined as the difference in flux between periods 1 and 2 after subtracting the Li75b flux.


2) Östergarnsholm
A similar test of the effects of drying the air for one Li-Cor 7200 instrument was also carried out at Östergarnsholm. In Fig. 7a the latent heat flux from Li72a without a Nafion dryer (blue) and Li72b with a Nafion dryer (red) is shown as a function of measured latent heat flux from a Li-Cor 7500 instrument. Only data from the open sea wind sector are used, and data with the same choice upon the variance of RSSI values (

A comparison of measured (a) latent heat flux and (b) CO2 flux between a Li-7500 instrument and two Li-7200 instruments at the Östergarnsholm site. A Nafion dryer was mounted on one of the Li-7200 instruments (Li72b). Data marked with dots denote a selection criterion for the variance of RSSI to be less than 0.001. Data with circles correspond to a stricter limit of about one-third of that. Data from Li72a without a Nafion dryer (blue) and data from Li72b with a Nafion dryer (red) are shown. Two fitted lines are included for the most high-quality data using the same color scheme, and the fitted expressions are given in the upper-left corner of each plot.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

A comparison of measured (a) latent heat flux and (b) CO2 flux between a Li-7500 instrument and two Li-7200 instruments at the Östergarnsholm site. A Nafion dryer was mounted on one of the Li-7200 instruments (Li72b). Data marked with dots denote a selection criterion for the variance of RSSI to be less than 0.001. Data with circles correspond to a stricter limit of about one-third of that. Data from Li72a without a Nafion dryer (blue) and data from Li72b with a Nafion dryer (red) are shown. Two fitted lines are included for the most high-quality data using the same color scheme, and the fitted expressions are given in the upper-left corner of each plot.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
A comparison of measured (a) latent heat flux and (b) CO2 flux between a Li-7500 instrument and two Li-7200 instruments at the Östergarnsholm site. A Nafion dryer was mounted on one of the Li-7200 instruments (Li72b). Data marked with dots denote a selection criterion for the variance of RSSI to be less than 0.001. Data with circles correspond to a stricter limit of about one-third of that. Data from Li72a without a Nafion dryer (blue) and data from Li72b with a Nafion dryer (red) are shown. Two fitted lines are included for the most high-quality data using the same color scheme, and the fitted expressions are given in the upper-left corner of each plot.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
The linear fits to the highest-quality data show a similar result as in Marsta with a systematic loss in latent heat flux for the enclosed-path Li72a instrument. This loss is here seen to be somewhat larger (of about 29%), but the range of measured fluxes is also recognized as lower, between −10 and 50 W m−2, at Östergarnsholm in comparison to Marsta, where it was between −20 and 190 W m−2. The number of data is also fewer (only 58 data of the highest quality), which limits the possibility to assess with certainty the amount of systematic loss of latent heat flux during this experiment. It is nevertheless clear that the results are qualitatively similar for Östergarnsholm and Marsta. The amount of flux loss may depend on a number of factors influencing the amount of covariance at short time scales. Some of these factors are wind speed, measurement height, and difference in surface roughness between Marsta and Östergarnsholm. From Fig. 7a it is also clear that the Nafion dryer worked well to remove water vapor flux. The latent heat flux was less than 4% for Li72b in comparison to the fluxes measured by the Li-Cor 7500 instrument.
The measured CO2 flux from Li-7500 was compared to the fluxes from the dried and undried Li-Cor 7200 instruments (Fig. 7b) with the same symbols and color scheme. The result without drying the air (Li72a) indicates systematic smaller CO2 fluxes of about 14% for the most high-quality data by the enclosed-path system. That result can be compared to the Li72b results, which include drying of the air, which show only a slight systematic reduction (less than 0.1%). Given the small number of data, it is difficult to clearly assess the amount of systematic reduction, or whether there is a reduction in CO2 fluxes at all.
To test the significance of our preliminary conclusions, we carried out a statistical test also for the Östergarnsholm data. In Table 3 a statistical summary of the experiment with mean CO2 fluxes of the three instruments is listed in column 1, which is noted to be small, on the order of 10−3 ppm m s−1. The mean flux values for the two Li-Cor 7200 instruments are also seen to be very similar, differing by less than 2% from each other. From Fig. 7b we also note that the variation in measured fluxes is between about −0.020 and 0.014 ppm m s−1 for the most high-quality Östergarnsholm data. This can be compared with period 2 in Marsta, where the range of data is between −0.6 and 0.25 ppm m s−1; thus, much more variable fluxes were observed at Marsta. The lower variability also made the data reasonably well approximated by a Gaussian distribution even without forming a difference time series by subtraction of a measured reference time series. Therefore, a two-sample t test could be considered here to directly test the change in the mean flux between the different instruments. In column 3 of Table 3, we listed both the difference in mean CO2 flux between the Li-Cor 7500 instrument and the undried Li72a, which is larger, and the 10-times-smaller difference between the dried and undried Li-Cor 7200 instruments. Neither of these differences was found by the t test to be statistically significant. The difference in mean flux value between dried and undried Li-Cor 7200 instruments was significant only at a 4% confidence level. The larger difference between the Li-Cor 7500 and the undried Li-Cor 7200 was significant at a confidence level of 49%, which we consider too low to be truly conclusive.
Statistical summary of the experimental field test with and without a Nafion dryer mounted on one of the Li-7200 instruments at the Östergarnsholm site. There were 58 high-quality data remaining when using the selection criterion


c. Will measured CO2 fluxes be affected by sea salt contamination on the windows of instruments?
During measurement period 4 in June 2015 at Marsta, some data were collected when the Li-Cor 7200 instruments (with and without Nafion dryer) were subjected to spraying of a saline solution on the windows of the instruments. This was used to test the hypothesis that hygroscopic deposits on the instrumental windows can lead to faulty and enhanced CO2 flux estimates. In that case, this effect could be part of a possible explanation for the sometimes reported order-of-magnitude difference of fluxes in comparison to typical bulk formulations for air–sea exchange.
First, we compared the measured CO2 fluxes during period 3 (Fig. 8a), which preceded the sea salt experiment and used a setup similar to the one in period 2. The main result is similar with less than 3% systematic difference between the two Li-7500 instruments and 5%–7% lower fluxes for the enclosed-path sensors in comparison to one of the Li-Cor 7500 instruments. The variation of measured fluxes for period 3 is larger in June than the other periods with 10-min flux values between −0.65 and 0.35 ppm m s−1. This is interpreted to be caused by environmental changes on the surrounding crop fields. Choosing a period close to the reference period when the sea salt experiment was conducted ensures that changing environmental factors would be minimal.

A comparison of measured CO2 fluxes between the Li75a instrument and the other three instruments under conditions (a) without salt and (b) with salt. A comparison of the two Li-7500 instruments is shown (black crosses). A comparison of fluxes between the undried Li-7200 without a mounted Nafion dryer and one of the Li-7500 instruments (blue circles). A comparison of the CO2 fluxes with the dried Li-7200 to the Li-7500 (red dots). Three fitted lines are included with the data in the corresponding color scheme, and the fitted expressions are given in the upper-left corner of each plot.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

A comparison of measured CO2 fluxes between the Li75a instrument and the other three instruments under conditions (a) without salt and (b) with salt. A comparison of the two Li-7500 instruments is shown (black crosses). A comparison of fluxes between the undried Li-7200 without a mounted Nafion dryer and one of the Li-7500 instruments (blue circles). A comparison of the CO2 fluxes with the dried Li-7200 to the Li-7500 (red dots). Three fitted lines are included with the data in the corresponding color scheme, and the fitted expressions are given in the upper-left corner of each plot.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
A comparison of measured CO2 fluxes between the Li75a instrument and the other three instruments under conditions (a) without salt and (b) with salt. A comparison of the two Li-7500 instruments is shown (black crosses). A comparison of fluxes between the undried Li-7200 without a mounted Nafion dryer and one of the Li-7500 instruments (blue circles). A comparison of the CO2 fluxes with the dried Li-7200 to the Li-7500 (red dots). Three fitted lines are included with the data in the corresponding color scheme, and the fitted expressions are given in the upper-left corner of each plot.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
Second, the measured CO2 fluxes from the sea salt experiment in period 4 are compared between instruments (Fig. 8b). The Li-Cor 7500 instruments compared very well for this limited dataset of 47 high-quality data points (with
Box plots of the difference in measured CO2 flux between one of the Li-Cor 7500 instruments and the other three instruments (Fig. 9) for period 3 and period 4 (which is denoted as the “salt” period) was used to illustrate the difference in behavior in the presence of salt. From the light gray boxes on the left, the two Li-Cor 7500 instruments are compared and a slight reduction in variance is observed for period 4, but this change is rather small and not very conclusive, as the number of data in period 4 is only 47 high-quality data compared to 428 10-min values for period 3. The median marked by the dashed–dotted line does also not change very much. For period 4 it is noted that some skewness is observed with more values less than the median value than above in the box (representing the 25th and 75th percentiles, respectively), but this is also not a very large change given the limited number of data in period 4. The white boxes show the result for the Li72b instrument with a Nafion dryer and here also the distributions do not change very much between the two periods. For the dark gray boxes on the right, which show the result for the undried Li72a instrument, the difference between the periods is larger with a change both in median value and a shift in the distribution. For period 3 the difference between the undried Li72a CO2 fluxes and the Li75a fluxes is somewhat skewed so that more values are positive because of the larger uptake values of CO2 measured at the Li-Cor 7500 instrument. In period 4 with hygroscopic deposits on the Li72a instrument windows, the distribution is shifted so that more negative values than positive values are observed because the higher uptake values of CO2 are measured by the enclosed-path sensor.

Boxplots of the difference in measured CO2 flux between the Li75a instrument and the other three instruments for two different periods without and with salt on the windows of the Li-7200 instruments. The light gray boxes show the comparison of the two Li-7500 instruments. The white boxes show the comparison of the Li-7200 instrument with a Nafion dryer installed (Li72b). The dark gray boxes show the comparison of the Li-7200 instrument without a Nafion dryer (Li72a). The median value is marked (dash–dotted line), and the boxes marks the 25th and 75th percentiles. Outliers are drawn as plus (+) signs if they are larger than Q3 + 1.5(Q3 − Q1); those smaller than Q1 − 1.5(Q3 − Q1) are drawn as whiskers of the data. Here Q1 and Q3 are the 25th and 75th percentiles, respectively. There are also a few outlier values for the period with no salt that lie outside the interval shown.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

Boxplots of the difference in measured CO2 flux between the Li75a instrument and the other three instruments for two different periods without and with salt on the windows of the Li-7200 instruments. The light gray boxes show the comparison of the two Li-7500 instruments. The white boxes show the comparison of the Li-7200 instrument with a Nafion dryer installed (Li72b). The dark gray boxes show the comparison of the Li-7200 instrument without a Nafion dryer (Li72a). The median value is marked (dash–dotted line), and the boxes marks the 25th and 75th percentiles. Outliers are drawn as plus (+) signs if they are larger than Q3 + 1.5(Q3 − Q1); those smaller than Q1 − 1.5(Q3 − Q1) are drawn as whiskers of the data. Here Q1 and Q3 are the 25th and 75th percentiles, respectively. There are also a few outlier values for the period with no salt that lie outside the interval shown.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
Boxplots of the difference in measured CO2 flux between the Li75a instrument and the other three instruments for two different periods without and with salt on the windows of the Li-7200 instruments. The light gray boxes show the comparison of the two Li-7500 instruments. The white boxes show the comparison of the Li-7200 instrument with a Nafion dryer installed (Li72b). The dark gray boxes show the comparison of the Li-7200 instrument without a Nafion dryer (Li72a). The median value is marked (dash–dotted line), and the boxes marks the 25th and 75th percentiles. Outliers are drawn as plus (+) signs if they are larger than Q3 + 1.5(Q3 − Q1); those smaller than Q1 − 1.5(Q3 − Q1) are drawn as whiskers of the data. Here Q1 and Q3 are the 25th and 75th percentiles, respectively. There are also a few outlier values for the period with no salt that lie outside the interval shown.
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
A statistical summary of our sea salt experiment (Table 4) corroborates the results about the effect of salt. It is directly seen from column 4 that the change in the mean flux behavior for Li72b with a Nafion dryer is only 0.0001 ppm m s−1, which is a 9-times-smaller change than between the two Li-Cor 7500 instruments (which was not subjected to salt contamination). For the Li72a instrument without drying of the air, the effect of salt contamination was much larger with a change of 0.0139 ppm m s−1. This is a factor-of-15.4-times-larger change than for the two Li-Cor 7500 instruments. Although we have some observed skewness in the distributions, a two-sample t test applied to these different situations was done. For the undried Li72a instrument, the change was then found statistically significant at a confidence level of 99% (column 5 in Table 4), providing a strong argument that salt contamination can lead to enhanced CO2 fluxes. The change between the two Li-Cor 7500 instruments for the two periods was smaller and statistically significant only at a low confidence level of 28%, indicating no real significance. The Nafion-dried Li72b also had only a very small change that was completely insignificant (significant only at a 3% confidence level). These levels of confidence should be interpreted carefully because of the observed skewness in the distributions and that the number of data is relatively few (<50), but they show that some issues of salt contamination can exist when measuring CO2 fluxes.
Statistical summary of the experimental field test with and without salt on the windows of Li-7200 instruments. In this field experiment from June 2015, there were 428 high-quality data (10-min values) available before a saline solution was sprayed on the windows and 47 high-quality data with salt-covered windows remained.


4. Summary and conclusions
To answer the question of whether drying the sampling air is important for measured CO2 fluxes, field data were collected and analyzed from one agricultural site, Marsta, and one marine site, Östergarnsholm. The conclusion from the agricultural site was that no important change of CO2 flux as a result of the drying of air was found. These conclusions are certain, as they are based on a fair amount of high-quality data (more than 1300 10-min values). There was, however, a systematic loss of flux in enclosed-path sensors of about 6%–7% for CO2 and 9% for the latent heat flux in comparison to two open-path sensors, which only differed about 1% for CO2 and 2% for latent heat fluxes. The underestimation of the enclosed-path sensors was partly from losses of flux at high frequencies (above 0.7 Hz), but some flux loss for CO2 was found to take place also over a broader range of turbulent scales for frequencies above 3/1000 Hz. Further analysis of spectra and cospectra in varying atmospheric conditions would be required to gain more insights but was considered beyond the scope of the present study.
At the marine site, qualitatively similar results were reached with undried and dried enclosed-path sensors differing by less than 2%, and no statistically significant difference (only 4%) was found for the mean CO2 flux value. The reduction in latent heat flux was found to be larger compared to the Marsta dataset, but the range of measured fluxes was smaller and the number of data points of the highest quality were fewer (<60). This limited the possibility to assess with certainty the amount of systematic flux loss from the Östergarnsholm dataset.
The recommendation for how to best measure CO2 fluxes needs in our opinion to be different for different environmental conditions. In situations without salt contamination, our study concludes no clear effect on measured CO2 fluxes from the presence of water vapor as long as instrument windows are kept clean. Thus, there is no need for extra dryer equipment in such situations. This eliminates the obvious drawback of placing a dryer for the intake airstream: that another instrument would then be needed to measure the latent heat flux. It also reduces the spectral attenuation that can be expected from using extra tubing.
According to our study, the salt contamination issue and the flux loss issue mainly at high frequencies are the two important issues that should be considered during the planning phase of an experiment. Cospectral flux loss is reduced using open-path sensors. In our study more high-quality data were retained when using open-path sensors in comparison to enclosed-path sensors. At sites where no obvious risk of optical contamination exists, a preliminary recommendation is to use open-path sensors for monitoring CO2 fluxes.
For our land-based study, we found an effect of the salt-promoted cross sensitivity that altered the CO2 fluxes by approximately 13%. This bias was statistically significant. On sites with small fluxes, the magnitude of absolute errors that this 13% change corresponds to is important, and several open ocean experiments (Miller et al. 2010; Landwehr et al. 2014; Blomquist et al. 2014) have shown order-of-magnitude differences. Using a Nafion dryer eliminated the influence from the effects of hygroscopic deposits on the instrument windows for our studied enclosed-path sensor, but the number of available data is admittedly low.
In the short-term application of infrared gas analyzers (IRGAs) in environments prone to sea salt, we support the conclusions from Miller et al. (2010); Landwehr et al. (2014) and Blomquist et al. (2014) that drying the air is to be recommended when measuring CO2 fluxes. We would for the future also recommend feasibility studies for new instruments to include self-cleaning mechanisms or other ways for removing hygroscopic deposits from instrument windows. This would be a preferred option in comparison to using dryer equipment, since from our study we conclude no effect on measured CO2 fluxes from water vapor itself. Only the combination of salt and water vapor showed a clear effect on measured CO2 fluxes. We recognize that currently with the available measurement technology, drying the intake air may be the best practical option for ocean flux measurement studies of CO2.
Regarding the issue of assessing the effects from water vapor and salt contamination separately, it may be important to reiterate a couple of things that make this study different from previous studies. A main difference between our Marsta data and studies conducted over open or coastal oceans is that at Marsta, cleaning away the salt is 100% effective. For measurement at sea, even the clean instrument is exposed to the salty air, so the “cleanliness” may last only some seconds or minutes. Optical contamination has been discussed in some studies to probably not be the source of cross sensitivity (Blomquist et al. 2014) but rather the water vapor itself. This can appear contradictory to our results. Blomquist et al. (2014) also mention, however, the possibility that relevant contaminants could be somewhat resistant to removal by rinsing. We also focused here on the measured CO2 fluxes rather than the mean CO2 concentration, which in Kondo et al. (2014) and Blomquist et al. (2014) was shown to be influenced by water vapor. Further study of sensitivity in results to both the amount of hygroscopic deposits on lenses and the resistance to removal of salt from instrumental windows in various conditions may be needed. The presented dataset and analysis suggest, however, that salt contamination can be a significant issue, leading to biased CO2 fluxes. Furthermore, water vapor itself without salt had no important effect on our measured CO2 fluxes.
Acknowledgments
We thank all members of the Meteorology group at Uppsala University for the discussions about this work. Frank Griessbaum, Gerardo Fratini, and George Burba from the Li-Cor company are also thanked for their constructive comments and discussions concerning this work during the GHG flux workshop “From Natural to Urban Systems” in Potsdam, Germany, in October 2015. We also thank two anonymous reviewers for their help with improving the manuscript. The sensor intercomparison was partly funded by ICOS Sweden and Vetenskapsrådet Grants 2012-03902 and 2013-02044. BW and SL were supported by the Norwegian Research Council under Projects 233901 and 244262, and by an award from the Irish Marine Institute under Grant PBA/ME/16/01.
APPENDIX
Exploring the Effect of Filter Methods and Thresholds
The squared correlation coefficient (r2) for CO2 fluxes between the Li-Cor 7200 and Li-Cor 7500 instruments in Figs. A1a and A1b show that the big drop in correlation during May occurred for the dried sensor Li72b (Fig. A1a). It did not occur for the undried enclosed-path instrument Li72a (Fig. A1b). This indicates the need for a strict threshold limit on

The squared correlation coefficient for the CO2 flux (black) and the fraction of remaining data (blue) are shown as a function of different threshold limits for
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

The squared correlation coefficient for the CO2 flux (black) and the fraction of remaining data (blue) are shown as a function of different threshold limits for
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
The squared correlation coefficient for the CO2 flux (black) and the fraction of remaining data (blue) are shown as a function of different threshold limits for
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
As an alternative way to obtain high-quality data, we tested choosing different threshold limits on RSSI values normalized by the median RSSI value (

The squared correlation coefficient for the CO2 flux (black) and the fraction of remaining data (blue) are shown as a function of different threshold limits for RSSI values normalized by the median RSSI value (
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

The squared correlation coefficient for the CO2 flux (black) and the fraction of remaining data (blue) are shown as a function of different threshold limits for RSSI values normalized by the median RSSI value (
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
The squared correlation coefficient for the CO2 flux (black) and the fraction of remaining data (blue) are shown as a function of different threshold limits for RSSI values normalized by the median RSSI value (
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
The squared correlation coefficient between CO2 fluxes measured by Li-Cor 7200b and Li-Cor 7500a (Fig. A3b) was high and equal (r2 = 0.98) for both methods. Comparing the Li-Cor 7200a and Li-Cor 7500a fluxes (Fig. A3a), the squared correlation coefficient was lower (r2 = 0.93) for the

The measured CO2 flux from Li-Cor 7500a (both April and May data are included) is compared to the measured fluxes from (a) Li-Cor 7200a, (b) Li-Cor 7200b, and (c) Li-Cor 7500b. Data for the chosen threshold limit
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

The measured CO2 flux from Li-Cor 7500a (both April and May data are included) is compared to the measured fluxes from (a) Li-Cor 7200a, (b) Li-Cor 7200b, and (c) Li-Cor 7500b. Data for the chosen threshold limit
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
The measured CO2 flux from Li-Cor 7500a (both April and May data are included) is compared to the measured fluxes from (a) Li-Cor 7200a, (b) Li-Cor 7200b, and (c) Li-Cor 7500b. Data for the chosen threshold limit
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
A looser limit gives more scatter with both outliers above and below a 1:1 line as shown by Fig. A4, where a selection criteria

The measured CO2 flux from Li-Cor 7500a (both April and May data are included) is compared to the measured fluxes from (a) Li-Cor 7200a, (b) Li-Cor 7200b, and (c) Li-Cor 7500b. Data for the chosen threshold limit
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1

The measured CO2 flux from Li-Cor 7500a (both April and May data are included) is compared to the measured fluxes from (a) Li-Cor 7200a, (b) Li-Cor 7200b, and (c) Li-Cor 7500b. Data for the chosen threshold limit
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
The measured CO2 flux from Li-Cor 7500a (both April and May data are included) is compared to the measured fluxes from (a) Li-Cor 7200a, (b) Li-Cor 7200b, and (c) Li-Cor 7500b. Data for the chosen threshold limit
Citation: Journal of Atmospheric and Oceanic Technology 35, 4; 10.1175/JTECH-D-17-0072.1
REFERENCES
Baldocchi, D., and Coauthors, 2001: FLUXNET: A new tool to study the temporal and spatial variability of ecosystem-scale carbon dioxide, water vapor, and energy flux densities. Bull. Amer. Meteor. Soc., 82, 2415–2434, https://doi.org/10.1175/1520-0477(2001)082<2415:FANTTS>2.3.CO;2.
Blomquist, B., B. Huebert, C. Fairall, L. Bariteau, J. Edson, J. Hare, and W. McGillis, 2014: Advances in air–sea CO2 flux measurement by eddy correlation. Bound.-Layer Meteor., 152, 245–276, https://doi.org/10.1007/s10546-014-9926-2.
Burba, G. G., D. K. McDermitt, A. Grelle, D. J. Anderson, and L. Xu, 2008: Addressing the influence of instrument surface heat exchange on the measurements of CO2 flux from open-path gas analyzers. Global Change Biol., 14, 1854–1876, https://doi.org/10.1111/j.1365-2486.2008.01606.x.
Grachev, A. A., L. Bariteau, C. W. Fairall, J. E. Hare, D. Helmig, J. Hueber, and E. K. Lang, 2011: Turbulent fluxes and transfer of trace gases from ship-based measurements during TexAQS 2006. J. Geophys. Res., 116, D13110, https://doi.org/10.1029/2010JD015502.
Halldin, S., and Coauthors, 1999: Continuous long-term measurements of soil–plant–atmosphere variables at an agricultural site. Agric. For. Meteor., 98–99, 75–102, https://doi.org/10.1016/S0168-1923(99)00149-5.
Hicks, B., and R. McMillen, 1988: On measurement of dry deposition using imperfect sensors and in non-ideal terrain. Bound.-Layer Meteor., 42, 79–84, https://doi.org/10.1007/BF00119876.
Högström, U., and Coauthors, 2008: Momentum fluxes and wind gradients in the marine boundary layer—A multi-platform study. Boreal Environ. Res., 13, 475–502.
Horst, T., 1997: A simple formula for attenuation of eddy fluxes measured with first-order-response scalar sensors. Bound.-Layer Meteor., 82, 219–233, https://doi.org/10.1023/A:1000229130034.
Horst, T., and D. Lenschow, 2009: Attenuation of scalar fluxes measured with spatially displaced sensors. Bound.-Layer Meteor., 130, 275–300, https://doi.org/10.1007/s10546-008-9348-0.
Kaimal, J. C., and J. J. Finnigan, 1994: Amospheric Boundary Layer Flows: Their Structure and Measurement. Oxford University Press, 289 pp.
Kljun, N., P. Calanca, M. W. Rotach, and H. P. Schmid, 2004: A simple parameterisation for flux footprint predictions. Bound.-Layer Meteor., 112, 503–523, https://doi.org/10.1023/B:BOUN.0000030653.71031.96.
Kohsiek, W., 2000: Water vapor cross-sensitivity of open path H2O/CO2 sensors. J. Atmos. Oceanic Technol., 17, 299–311, https://doi.org/10.1175/1520-0426(2000)017<0299:WVCSOO>2.0.CO;2.
Kondo, F., K. Ono, M. Mano, A. Miyata, and O. Tsukamoto, 2014: Experimental evaluation of water vapour cross-sensitivity for accurate eddy covariance measurement of CO2 flux using open-path CO2/H2O gas analysers. Tellus, 66B, 23803, https://doi.org/10.3402/tellusb.v66.23803.
Landwehr, S., S. D. Miller, M. J. Smith, E. S. Saltzman, and B. Ward, 2014: Analysis of the PKT correction for direct CO2 flux measurements over the ocean. Atmos. Chem. Phys., 14, 3361–3372, https://doi.org/10.5194/acp-14-3361-2014.
Landwehr, S., N. O. Sullivan, and B. Ward, 2015: Direct flux measurements from mobile platforms at sea: Motion and airflow distortion corrections revisited. J. Atmos. Oceanic Technol., 32, 1163–1178, https://doi.org/10.1175/JTECH-D-14-00137.1.
Lee, X., W. Massman, and B. Law, Eds., 2005: Handbook of Micrometeorology: A Guide for Surface Flux Measurement and Analysis. Atmospheric and Oceanographic Sciences Library, Vol. 29, Springer, 250 pp., https://doi.org/10.1007/1-4020-2265-4.
Massman, W., 2000: A simple method for estimating frequency response corrections for eddy covariance systems. Agric. For. Meteor., 104, 185–198, https://doi.org/10.1016/S0168-1923(00)00164-7.
McGillis, W. R., and Coauthors, 2004: Air-sea CO2 exchange in the equatorial Pacific. J. Geophys. Res., 109, C08S02, https://doi.org/10.1029/2003JC002256.
Miller, S., C. Marandino, and E. S. Saltzman, 2010: Ship-based measurement of air-sea CO2 exchange by eddy covariance. J. Geophys. Res., 115, D02304, https://doi.org/10.1029/2009JD012193.
Moore, C., 1986: Frequency response corrections for eddy correlation systems. Bound.-Layer Meteor., 37, 17–35, https://doi.org/10.1007/BF00122754.
Nakai, T., M. van der Molen, J. Gash, and Y. Kodama, 2006: Correction of sonic anemometer angle of attack errors. Agric. For. Meteor., 136, 19–30, https://doi.org/10.1016/j.agrformet.2006.01.006.
Nilsson, E. O., A. Rutgersson, and P. Sullivan, 2010: Flux attenuation due to sensor displacement over sea. J. Atmos. Oceanic Technol., 27, 856–868, https://doi.org/10.1175/2010JTECHA1388.1.
Prytherch, J., M. J. Yelland, R. Pascal, B. Moat, I. Skjelvan, and C. Neill, 2010: Direct measurements of the CO2 flux over the ocean: Development of a novel method. Geophys. Res. Lett., 37, L03607, https://doi.org/10.1029/2009GL041482.
Prytherch, J., M. J. Yelland, I. M. Brooks, D. J. Tupman, R. W. Pascal, B. I. Moat, and S. J. Norris, 2015: Motion-correlated flow distortion and wave-induced biases in air–sea flux measurements from ships. Atmos. Chem. Phys., 15, 10 619–10 629, https://doi.org/10.5194/acp-15-10619-2015.
Rutgersson, A., M. Norman, B. Schneider, H. Pettersson, and E. Sahlée, 2008: The annual cycle of carbon dioxide and parameters influencing the air–sea carbon exchange in the Baltic Proper. J. Mar. Syst., 74, 381–394, https://doi.org/10.1016/j.jmarsys.2008.02.005.
Rutgersson, A., M. Norman, and G. Åström, 2009: Atmospheric CO2 variation over the Baltic Sea and the impact on air–sea exchange. Boreal Environ. Res., 14, 238–249.
Sahlée, E., A.-S. Smedman, A. Rutgersson, and U. Högström, 2008: Spectra of CO2 and water vapour in the marine atmospheric surface layer. Bound.-Layer Meteor., 126, 279–295, https://doi.org/10.1007/s10546-007-9230-5.
Starkenburg, D., S. Metzger, G. J. Fochesatto, J. G. Alfieri, R. Gens, A. Prakash, and J. Cristóbal, 2016: Assessment of despiking methods for turbulence data in micrometeorology. J. Atmos. Oceanic Technol., 33, 2001–2013, https://doi.org/10.1175/JTECH-D-15-0154.1.
Vesala, T., and Coauthors, 2008: Flux and concentration footprint modelling: State of the art. Environ. Pollut., 152, 653–666, https://doi.org/10.1016/j.envpol.2007.06.070.
Wang, W., and Coauthors, 2016: Performance evaluation of an integrated open-path eddy covariance system in a cold desert environment. J. Atmos. Oceanic Technol., 33, 2385–2399, https://doi.org/10.1175/JTECH-D-15-0149.1.
Webb, E. K., G. I. Pearman, and R. Leuning, 1980: Correction of flux measurements for density effects due to heat and water vapour transfer. Quart. J. Roy. Meteor. Soc., 106, 85–100, https://doi.org/10.1002/qj.49710644707.